Dushyant Patel LC-MS/MS Method
S.K.P.C.P.E.R., Kherva 63 M .Pharm. Thesis
spiked calibration standards, LQC and HQC samples. Concentrations were calculated to
determine % mean change during stability period.
¾ GBP was found to be stable in LQC and HQC samples for 20 Hrs at room temperature
with % mean change of 4.17 and 3.71, respectively (Table No. 4.26).
TABLE NO: 4.26
BENCH TOP STABILITY OF GBP (20 hrs at room temp.)
Sr. No.
LQC (0.15 µg/ml)
Comparison samples Stability samples
Sample conc. % Nominal Conc. Sample conc. % Nominal Conc.
1 0.147 98.00 0.160 106.7
2 0.149 99.33 0.151 100.7
3 0.149 99.33 0.156 104.0
4 0.145 96.67 0.156 104.0
5 0.151 100.7 0.157 104.7
6 0.146 97.33 0.144 96.00
n 6
6
Mean 0.148 0.154
% Mean Change 4.17
Sr. No.
HQC (4.50 µg/ml)
Comparison samples Stability samples
Sample conc. % Nominal Conc. Sample conc. % Nominal Conc.
1 4.477 99.49 4.550 101.1
2 4.535 100.8 4.639 103.1
3 4.458 99.07 4.724 104.9
4 4.587 101.9 4.896 108.8
5 4.745 105.4 4.935 109.7
6 4.655 103.4 4.733 105.2
n 6
6
Mean 4.576 4.746
% Mean Change 3.71
4.3.10.3 Process stability of GBP (at 5°C in auto sampler for 76 hrs):
¾ LQC and HQC samples were prepared and processed. These processed samples were
kept in auto sampler for 76 hrs at 5°C, these samples were analyzed along with
previously bulk spiked calibration standards, LQC and HQC samples concentrations
were calculated to determine % mean change during stability period.
¾ GBP was found to be stable in LQC and HQC samples for 76 hrs at 5°C in auto sampler
with % mean change of 0.77 and 5.44, respectively (Table No. 4.27).