Dushyant Patel LC-MS/MS Method
S.K.P.C.P.E.R., Kherva 40 M .Pharm. Thesis
4.2.12 SAMPLE TREATMENT/SAMPLE PREPARATION:
¾ The study samples along with calibration standards and QC samples were taken out from
the -20°C/-70°C deep freezer and kept at room temperature for thawing. Vortexed each
sample adequately using a vortex mixer before pipetting. Followed the procedure from a
to d as given below.
a) Internal standard spiking solution (25µl) was transferred in micro tube except in blank
Collision energy : 12 V (For GBP), 15 V (For LEV)
Collision gas pressure : 1.5 mTorr
MS acquire time : 3.00 min.
No of segments : 1
Segment time : 3.00 min.
Chrom filter : 10 sec
Peak width : 0.01 for GBP & LEV.
No of scan events : 2
Scan event – 1 (For GBP)
Polarity : Positive
Data type : Centroid
Scan type : SRM (Selected Reaction monitoring)
Parent Product Width Time CE Q1PW Q3PW
172.15 154.15 0.010 0.200 12 0.70 0.70
Scan event - 2 (For LEV)
Polarity : Positive
Data type : Centroid
Scan type : SRM (Selected Reaction monitoring)
Parent Product Width Time CE Q1PW Q3PW
171.13 126.06 0.010 0.200 15 0.70 0.70
Where,
Parent : Parent ion
Product : Daughter ion being monitored
Width : Scan width
Time : Scan time for event
CE : Collision energy
Q1PW : Resolution for first quadrupole
Q3PW : Resolution for third quadrupole
Divert Valve Settings : Not in use.