¾ Unextracted standard solution of an analyte (at LQC, MQC and HQC level) and internal
standard of concentrations equivalent to concentration of extracted samples were
processed and analyzed.
¾ The area for extracted LQC, MQC, HQC samples and internal standard were compared
with areas of unextracted standards and internal standard that represent 100 % recovery.
¾ % recovery of an analyte at LQC, MQC and HQC samples and an internal standard were
calculated.
¾ % recovery for extracted HQC, MQC and LQC samples should not exceed 115%, if it
exceeds then reporting should be justified.
Acceptance criteria:
¾ Recovery of analyte and internal standard should be consistent, precise and reproducible.
¾ Variability within areas at each QC level for analyte should be within CV of 15%.
¾ Variability of the analyte recoveries across all the three QC levels should be within C.V
of 20%.
¾ Variability within areas of IS found with each extracted QC level should be within CV of
15%, while the variability of internal standard area found between all the QC levels
should be within CV of 20%.
H. Dilution integrity:
Procedure:
¾ The stock solutions of analyte and internal standard were prepared as per the procedure
described in sample preparation section.
¾ Analyte spiking stock solution was spiked in blank plasma to get concentration
equivalent to 2 times of ULOQ.
¾ Above spiked samples (2xULOQ) was diluted with blank plasma to get 1/2 and 1/4
concentrations of the spiked sample or as per requirement.
¾ Calibration standards and six aliquots each of diluted samples (1/2 and 1/4 dilutions)
were processed and analyzed as per the procedure as described in sample preparation
section.
¾ %CV for the back calculated concentration was calculated.
¾ % accuracy for the back calculated concentration was calculated.
Acceptance criteria:
¾ Back calculated concentration accuracy should be within 85-115% of theoretical
concentration.
¾ Precision (% CV) for the back calculated concentration should be within 15%.