7-18
Operation Manual
Storage SR Server Information
Select the ‘STORAGE SR’ item from the combo boxes at the top right corner. Set up the DICOM server to
transmit the SR le to.
Transfer Mode
X
After acquisition: Send an image whenever you press the Store button to save it.
X
On end exam: Send all saved images when you press the End Exam button.
X
Manually: Send an image the user selects from the Exam List or in Review.
Changing DICOM Information
Select a service from the list, and press Edit. Note that if any of the essential information for the service
(AE Title, IP Address, and Port) is missing, the service will be deactivated. After entering the information,
press Apply to apply the new information. Press Cancel to cancel.
Deleting a DICOM Service
Select a service on the screen, and press Delete. If any of the essential information for the service (AE
Title, IP Address, Port) is missing, the service will be deactivated.
Testing a DICOM Server
Select a service on the screen, and press Verify. Test the connection status of the selected services and
display the result of test in Ping and Verify elds each. Success means that the system has a working
connection.
NOTE: Depending on your organization’s network security settings, the Ping result may always be
displayed as Fail. Please contact your network administrator for more information.