Chapter 5 Measurements and Calculations
The equations used for D Trace measurement are as follows:
X
X
X
X
NOTE: If Application under Utility > Measure Setup > General > Caliper is set to Cardiac, results
for Vmax, Vmean, PGmax, PGmean, VTI, PHT, Acc, AccT, Dec, DecT, etc. are shown on the screen.
D Time
This is a basic measurement that is only available in Spectral Doppler Mode. You can specify two bars
on a Spectral Doppler image and calculate the time between them.
1. Press Caliper and select D Time under soft menu [1]. Time is shown in the user information area.
2. Set two bars on the spectrum with the trackball and the Set button.
3. When the measurement is nished, its result is shown on the screen.