4-4
Operation Manual
Basic Operating Instructions
The items that are commonly used in each diagnosis mode are shown below:
Using the Control Panel
The items that can be used in each diagnosis mode are provided as menu items. You can change the
image format or optimize an image to facilitate your diagnosis.
Q Scan/Gain
Q Scan: Pressing the relevant dial-button enables Quick Scan. The mark will appear at the top
of an image.
In 2D Mode, it is used to optimize the contrast and brightness of an image by adjusting Gain and
TGC automatically. In PW Spectral Doppler Mode, it is used to optimize the spectrum by adjusting
Scale and Baseline automatically.
NOTE: The Quick Scan function is only available with specific probes and applications.
X
Gain: Rotate this dial-button to adjust Gain. The Gain button appears dierently depending on
the diagnosis mode, but it is usually in the form of a dial-button used to select a diagnosis mode.
You can adjust the brightness of an image. If you rotate the Gain dial-button clockwise, its value
increases.
Angle
NOTE: Available in 2D mode and Spectral Doppler mode only.
Rotating the dial in Spectral Doppler mode adjusts the angle of the sample volume in 1° increments.
In 2D mode, rotating the dial scrolls the image left and right if Scan Area is not at 100%.
TGC (Time Gain Compensation)
Use the TGC sliders on the control panel.
In general, ultrasound penetration gets weaker with depth. TGC can be used to compensate for this
eect.