Samsung SONOACE R7 Uživatelský manuál Strana 28

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 320
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 27
1-12
Operation Manual
Approved Cables, Transducers and Accessories for EMC
Approved Cable for ElectroMagnetic Compliance
Cables connected to this product may aect its emissions; Use only the cable types and lengths listed
below table.
Cable Type Length
VGA Shielded Normal
Parallel Shielded Normal
RS232C Shielded Normal
USB Shielded Normal
LAN (RJ45) Twisted pair Any
S-Video Shielded Normal
Foot Switch Shielded 2.73yd
B/W Printer Unshielded Coaxial Normal
MIC Unshielded Any
Printer Remote Unshielded Any
Audio R.L Shielded Normal
VHS Shielded Normal
ECG AUX input Shielded < 3m
e-Motion Marker Shielded 3m
Approved Transducer for ElectroMagnetic Compliance
The probe listed in Chapter 9. Probes’ when used with this product, have been tested to comply with the
group1 class B emission as required by International Standard CISPR 11.
Approved Accessories for ElectroMagnetic Compliance
Accessories used with this product may eect its emissions.
CAUTION: When connecting other customer-supplied accessories to the system, such as a remote
printer or VCR, it is the users responsibility to ensure the electromagnetic compatibility of the
system. Use only CISPR 11 or CISPR 22, CLASS B compliant devices.
Zobrazit stránku 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 319 320

Komentáře k této Příručce

Žádné komentáře