Samsung SONOACE R3 Uživatelský manuál Strana 138

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 442
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 137
User Manual
5-14
1. Press the Sofmenu button [1] Distance to select Doppler Velocity (A/B). “Doppler Velocity (A/B)” is dis-
played on the left side of the screen.
2. Specify two points to measure velocity for.
Place the cursor at a desired position with the trackball, and press the Set button.
3. When the measurement is nished, its result is shown on the screen.
V_A: Velocity at Point A
V_B: Velocity at Point B
RI: Resistivity Index
A/B: The Ratio of Velocity
Doppler Trace
This is a basic measurement that is available in Spectral Doppler Mode only. You can specify a point in a
Spectral Doppler image and trace a curve from that point to calculate the velocity, integral value and aver-
age velocity of blood ow.
1. Press the Sofmenu button [1] Distance to select Doppler Trace. “Doppler Trace is displayed on the
left side of the screen.
2. Trace a curve. The method for measuring a curve is the same as in “2D Distance (Trace).
3. When the measurement is nished, its result is shown on the screen.
dT: Change in Time
VTI: Velocity Time Integral
Mean V: Mean Velocity
Peak V: Peak Velocity
Mean G: Mean Gradient
The equations used for D Trace measurement are as follows: