Chapter 5 Measurements and Calculations
Measurement Accuracy
Measurement values can vary, depending on the nature of the ultrasound, the body’s response to
ultrasound, the measurement tools, algorithms, product settings, probe type and user operation.
Before using this product, make sure to read and understand the following information regarding the
causes of measurement errors, and measurement optimization.
Causes of Measurement Errors
Image Resolution
The resolution of ultrasound images may be limited by the available space.
X
Errors due to a signal range may be minimized by adjusting focus settings. Optimizing focus
settings increases the resolution of the measurement area.
X
In general, lateral resolution is lower than axial resolution. Therefore, measurements should be
performed along the axis of the ultrasound beam to obtain accurate values.
X
Gain has a direct impact on resolution. Gain can be adjusted by using the Gain button for each
mode.
X
In general, increasing the frequency of ultrasound enhances resolution.
Pixel Size
X
An ultrasound images in the product consist of pixels.
X
Since a single pixel represents the basic unit of an image, a measurement error may result in
the displacement of approximately ±1 pixel when compared to the original image size.
X
However, this error becomes significant only when a narrow area in an image is measured.