4-125
Chapter 4 Diagnosis Modes
[Figure 4.65 Volume Slice – Touch Screen]
VolumeNT
NOTE: VolumeNT is an optional feature of this product.
Use the VolumeNT feature to locate the mid-sagittal view and measure the nuchal translucency (NT)
of the fetus. The VolumeNT window appears on the touch screen. Use the trackball to position the NT
Seed in the NT area and press the Set button to display the NT measurement in plane A.
Tips!
How to get good results
X
Obtain the fetuss sagittal view as precisely as possible.
X
A higher contrast between the fetuss palate and nasal bone will yield better results.
X
Preferably, the lateral orientation of the probe should be parallel to the orientation of the fetus’s
body.
X
For optimal results, make sure that the angle between the probe and the fetuss nasal bone is as
close to 30 degrees as possible.
Display Format
Press the touch screen and select a button to choose an image display method. When Full Screen is
selected, the image in plane A will be displayed in full screen mode.