Chapter 1 Safety
Additional equipment connected to medical electrical equipment must comply with the respective IEC or
ISO standards (e.g. IEC 60950 for data processing equipment). Furthermore all congurations shall comply
with the requirements for medical electrical systems (see IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-
1, respectively). Anybody connecting additional equipment to medical electrical equipment configures a
medical system and is therefore responsible that the system complies with the requirements for medical
electrical systems. Attention is drawn to the fact that local laws take priority over the above-mentioned
requirements. If in doubt, consult your local representative or the technical service department.
WARNING:
X
Electric shock may exist result if this system, including and all of its externally mounted recording and
monitoring devices, is not properly grounded.
X
Do not remove the covers on the system; hazardous voltages are present inside. Cabinet panels must
be in place while the system is in use. All internal adjustments and replacements must be made by a
qualified Samsung Medison Customer Service Department.
X
Check the face, housing, and cable before use. Do not use and disconnect the power source, if the face
is cracked, chipped, or torn, the housing is damaged, or if the cable is abraded.
X
Always disconnect the system from the wall outlet prior to cleaning the system.
X
All patient contact devices, such as probes and ECG leads, must be removed from the patient prior
to application of a high voltage defibrillation pulse.
X
The use of flammable anesthetic gas or oxidizing gases (N
2
O) should be avoided.
X
Avoid places where the system is likely to be difficult to operate the disconnection device.
X
Do not use HF surgical equipment with the system. Any malfunctions in the HF surgical
equipment may result in burns to the patient.
CAUTION:
X
The system has been designed for 100-120VAC and 200-240VAC; you should select the input
voltage of printer and VCR. Prior to connecting a peripheral power cord, verify that the voltage
indicated on the power cord matches the voltage rating of the peripheral device
X
An isolation transformer protects the system from power surges. The isolation transformer
continues to operate when the system is in standby.
X
Do no
t immerse the cable in liquids. Cables are not waterproof.
X
The auxiliary socket outlets installed on this system are rated 100-120V and 200-240V with
maximum total load of 200W. Use these outlets only for supplying power to equipment that is
intended to be part of the ultrasound system. Do not connect additional multiple-socket outlets
or extension cords to the system.
X
Do not connect peripheral devices, not listed in this manual, to the auxiliary socket outlets of the
system.
X
Do not touch SIP/SOP and the patient simultaneously. There is a risk of electric shock from leakage
current.
Zobrazit stránku 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 267 268

Komentáře k této Příručce

Prof Akhigbe 24 Apr 2024 | 14:29:06

I want to learn how to do angle correction on PW Doppler on my machine.

Prof Akhigbe 24 Apr 2024 | 14:30:30

How do I assess PW angle cursor?

Prof Akhigbe 24 Apr 2024 | 14:40:21

I don't want to join the group.