6-13
Chapter 6 Image Management
e-Motion Marker (Optional)
e-Motion Marker displays the probe's position and angle on the screen at the location being scanned.
The e-Motion Marker Accessory mounted on the probe detects positional data, which is rendered as a
3D image.
[Figure 6.7 e-Motion Marker Uterus+Axis Model]
NOTE:
X
The e-Motion Marker Accessory is needed to use this option.
X
The e-Motion Marker can only be used if the following conditions are met.
Probe: EVN4-9
Application: Gynecology
Preset: Uterus
X
Refer to the installation guide that comes with the e-Motion Marker package for instructions on
installing e-Motion Marker.
X
BodyMarkers cannot be used while e-Motion Marker is being used.
WARNING:
X
The e-Motion Marker provides reference information to assist in diagnosis. It does not provide
any numerical diagrams or final diagnosis.
X
The e-Motion Marker displays the relative directions of a beam plane, so the direction displayed
may differ from the actual probe direction because of accumulated errors. Therefore, press the
Initialize button to initialize the position before starting to use the e-Motion Marker.
X
e-Motion Markers circuitry contains a sensor, and must not be dropped or subjected to other
forms of mechanical impact. Performance may degrade, and damage to the exterior, such as
breakage or cracking, may occur.
X
The e-Motion Marker meets the IPX 7 rating. However, do not cleanse or sterilize the probe while
the e-Motion Marker is mounted.
X
Do not use e-Motion Marker near metallic objects, magnetic objects, or products that contain
high electric current, because they may affect the magnetic sensor and cause malfunction.