8-8
Operation Manual
Accuracy Check
NOTE: The user must ensure that safety inspections are performed every 2 years according to the
requirements of safety standard EN 60601-1. Only trained persons are allowed to perform these
safety inspections.
The product’s maintenance status may aect the measurements obtained using the product. The product
should be maintained in an optimal state to ensure reliable measurements.
To ensure optimal operation of the product, perform an accuracy check every year. The equations and
table related to measurement accuracy are included in “Chapter 5 Measurements in this manual.