Operation Manual
1-10
EMC
The testing for EMC(ElectroMagnetic Compatibility) of this system has been performed according to the
international standard for EMC with medical devices (IEC60601-1-2). This IEC standard was adopted in Europe
as the European norm (EN60601-1-2).
Guidance and manufacturer’s declaration - electromagnetic emission
This product is intended for use in the electromagnetic environment specied below. The customer or the
user of this product should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment -guidance
RF Emission
CISPR 11
Group 1
The Ultrasound System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emission
CISPR 11
Class B
The Ultrasound System is suitable for use in all establishments,
other than domestic and those directly connected to the public
low-voltage power supply network that supplies building used for
domestic purpose.
Harmonic Emission
IEC 61000-3-2
Class A
Flicker Emission
IEC 61000-3-3
Complies